Using Instrument Functions
Using Markers
The three marker math functions: statistics, flatness, and RF filter stats,
perform certain mathematical calculations on the amplitude data of
user-defined trace segments.
For measurement channel 1, the trace segment is dehned with markers 1 and
2; for measurement channel 2, the trace segment is
with markers 3
and 4. The marker math parameters are updated after each sweep, as well as
any time a marker is moved. Regular marker tracking is not available with
the marker math functions.
To Use Marker
Statistics
N O T E
You cannot have marker math functions active at the same time as marker search functions.
The marker statistics function measures a user-defined segment of the
measurement trace and calculates the following:
frequency span
l
mean amplitude
l
standard deviation of the amplitude response
peak-to-peak ripple
Limit testing may be performed on the statistical mean and peak-to-peak
ripple. See “‘lb Use Marker Limit Functions, later in this chapter for
information.
1. On measurement channel 1 press
and set markers 1 and 2
to
the beginning and end of the trace segment that you want to
measure. (When using measurement channel 2, use markers 3 and 4 to
define the trace segment.)
Press
Nath Statistics.
4-20
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Содержание 8711C
Страница 18: ...I 1 Installing the Analyzer 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 103: ...4 Using Instrument Functions 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 203: ...5 Optimizing Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 258: ...7 Automating Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 452: ...I 10 Specifications and Characteristics 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt co ...
Страница 473: ...11 Safety and Regulatory Information 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 480: ...12 Preset State and Memory Allocation 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...