I
Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
This method of calibration is only necessary when trying to achieve
maximum dynamic range 100
A response and isolation calibration
prompts you to connect loads to both ports and then to connect a through
cable. These standards are measured across the frequency band you have
defined, using the number of points you have defined. These measurements
are used to remove systematic errors caused by frequency response and
crosstalk.
This method of calibration is superior to the standard response calibration.
This calibration prompts you to connect four measurement standards: an
open, a short, a load, and a through cable. The analyzer measures each
standard across the frequency band you have
using the number of
points you have defined. The measurements of these standards are used to
remove systematic errors caused by frequency response and source match.
Transmission
Calibration
Interpolation
Widening the frequency span after performing any of these calibrations will
invalidate them, and restore the default calibration. You may narrow the span
and the analyzer
interpolate correction for the narrower span.
Note that after you have calibrated, a
appears in the upper right hand
corner of the display. This “C” indicates that a user-defined
is in use. If
you change to a narrower span, note that the
changes to “C?“, indicating
the analyzer is now interpolating between calibrated measurement points.
The “C?” notation also appears when other system parameters, such as
power, number of points, or sweep time, have changed.
6-12
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Содержание 8711C
Страница 18: ...I 1 Installing the Analyzer 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 103: ...4 Using Instrument Functions 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 203: ...5 Optimizing Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 258: ...7 Automating Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 452: ...I 10 Specifications and Characteristics 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt co ...
Страница 473: ...11 Safety and Regulatory Information 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 480: ...12 Preset State and Memory Allocation 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...