I
Calibrating for Increased Measurement
Accuracy
Check the Calibration
3. Plot the results to disk or to hard copy for later comparison.
4. Compare subsequent calibration checks against the initial one and keep
track of the tune interval between the first calibration check and later
ones.
5. When the residual errors have drifted beyond what is acceptable, note
the tune interval and use that information to determine what is the best
calibration interval for your measurements.
6. Repeat the process above enough times to feel confident that your
calibration interval is correct.
The calibration check feature can be used on any of the transmission or
reflection calibrations available (including the factory default calibration). See
6-2.
Table 6-2. Calibration Check Error Terms
Measurement
Calibration
Standards’
Error
Needed
Computed
Transmission
Default
Source Match, load Match, Transmission Tracking
Response
Source Match, load Match, Transmission Tracking
Response Isolation
Source Match, load Match, Transmission Tracking,
Enhanced Response
Source Match, bad Match, Transmission Tracking
Test
Source Match, Load Match, Transmission Tracking
Reflection
Default
Diractivity, Source Match, Reflection Tracking
One Port
Diractivity, Source Match, Reflection Tracking
Test
Diractivity, Source Match, Reflection Tracking
1
1 Open, Short, L Load, T = Thru
2 See “Error Term Descriptions and Typical Values”, later in this chapter,
3 Used with HP
multiport test set only
6-27
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com
Содержание 8711C
Страница 18: ...I 1 Installing the Analyzer 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 36: ...I I 2 Getting Started 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 55: ...3 Making Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 99: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 103: ...4 Using Instrument Functions 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 203: ...5 Optimizing Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 258: ...7 Automating Measurements 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 276: ...深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 340: ...I 8 Front Rear Panel 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 364: ...9 Softkey Reference 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt c ...
Страница 452: ...I 10 Specifications and Characteristics 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt co ...
Страница 473: ...11 Safety and Regulatory Information 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 480: ...12 Preset State and Memory Allocation 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...
Страница 499: ...Index I 深圳市盛腾仪器仪表有限公司 Tel 0755 83589391 Fax 0755 83539691 Website www Sengt com ...