Triggering the HP 54645A Oscilloscope
The HP 54645A Oscilloscope allows you to synchronize the display to
the actions of the circuit under test by defining a trigger condition.
The triggering modes include Auto Lvl, Auto, and Normal. Triggering
types include Glitch, TV, and Slope/Edge and you can use channels 1
or 2, Line, or the External Trigger BNC for the source.
MegaZoom Technology Simplifies Triggering
With the built-in MegaZoom technology, you can simply Autoscale the
waveforms, then stop the scope to capture a waveform. You can then pan and
zoom through the data to find a stable trigger point. Autoscale often produces a
triggered display.
To use external triggering
The external trigger input is only available on the HP 54645A Oscilloscope.
1
Connect a signal to the BNC labeled “External Trigger”.
The external trigger input has an input impedance of 1 M
Ω
, and can accept a
maximum signal of
±
400 V (dc + peak ac) with a range of
±
18 V. You can
connect the trigger output of other instruments to this external trigger input,
or use an oscilloscope probe, such as the HP 10074A.
2
Press
Slope/Glitch
. Then select the rising or falling edge, or
Glitch
.
If TV is selected, you cannot use the external trigger, and you must select
either channel 1 or channel 2. If you select glitch, define the polarity and
duration qualifier for the glitch.
3
Press
Source
. Then select
Ext
.
The external trigger is always dc coupled. If you select ac coupling or
low-frequency reject, these functions are not active until you change the
trigger source to channel 1, channel 2, or line. Notice that the status line and
softkeys reflect this.
3-7
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...