![HP 54645A Скачать руководство пользователя страница 209](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177209.webp)
1
Connect the equipment.
a
Connect the signal generator to the input of the power splitter.
b
Connect the power sensor to one output of the power splitter, and
connect channel 1 of the oscilloscope to the other power splitter
output. Put a 50-
Ω
feedthrough at the input of the oscilloscope.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
2
Set up the oscilloscope.
a
Press
Setup
, then press the
Default Setup
softkey.
b
Set the time base to 500 ns/div.
c
Set the Volts/Div for channel 1 to 200 mV/div.
d
Press
Display
, then press the
Average
softkey.
e
Toggle the
# Average
softkey to select
8
averages.
3
Set the signal generator for 1 MHz at about +11.6 dBm.
Notice that the signal on the display is about 5 cycles and six divisions of
amplitude.
4
Press
Voltage
, then press the
Vp-p
softkey.
Wait a few seconds for the measurement to settle (averaging is complete),
then note the V
p-p
reading from the bottom of the display.
V
p-p
= _______ V.
5
Set the calibration factor percent of the power meter to the 1 MHz
value from the calibration chart on the probe, then press dB (REF)
on the power meter to set a 0 dB reference.
6
Change the frequency of the signal generator to 100 MHz.
7
Set the calibration factor of the power meter to the 100 MHz percent
value from the chart on the probe (interpolate the value between two
frequencies, if necessary).
Adjust the amplitude of the signal generator for a power reading as close as
possible to 0.0 dB (REL). Power meter reading = ______ dB.
Testing, Adjusting, and Troubleshooting
To verify bandwidth
7-20
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...