![HP 54645A Скачать руководство пользователя страница 185](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177185.webp)
To trigger on odd or even fields
To check the envelope of your video signals, or to measure worst case
distortion, trigger on the odd or even fields. When Field 1 is selected, the
oscilloscope triggers on color fields 1 or 3. When Field 2 is selected, the
oscilloscope triggers on color fields 2 or 4.
1
Select the TV display, TV as the trigger mode, and the appropriate TV
standard as described in the section, "To autoscale on a video signal."
2
Press
Slope/Glitch
in the
TRIGGER
section of the front panel,
then press the
TV
and
TV Menu
softkeys. Press
Field 1
or
Field 2
.
The trigger circuits look for the position of the start of Vertical Sync to
determine the field. But this definition of field does not take into
consideration the phase of the reference subcarrier. When Field 1 is selected,
the trigger system will find any field where the vertical sync starts on Line 4.
In the case of NTSC video, the oscilloscope will trigger on color field 1
alternating with color field 3 (see the following figure). This setup can be
used to measure the envelope of the reference burst.
Triggering on Color Field 1 Alternating with Color Field 3
Figure 6-5
Using Option 005 Enhanced TV/Video Trigger (HP 54645A)
To trigger on odd or even fields
6-10
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...