6
Adjust the channel 1 low-frequency compensation adjustment for as
flat a pulse top as possible.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
7
Repeat steps 1 through 6 for channel 2.
Channel 2 is labeled "2" on HP 54645A and "A2" on HP 54645D.
Variable Capacitor Reference Designators
To adjust the high-frequency pulse response
In this procedure, you adjust the high-frequency pulse response for each
channel.
Table 7-15
Equipment Required
Equipment
Critical specifications
Recommended Model/Part
Pulse generator
Rise time <875 ps
PSPL 1107B TD and
PSPL 1110B Driver
Adapter
SMA (f) to BNC (m)
HP 1250-1787
Feedthrough
50
Ω
, BNC (m) and (f)
HP 11048C
Figure 7-6
Testing, Adjusting, and Troubleshooting
To adjust the high-frequency pulse response
7-36
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...