5
Set the time base to 200
µ
s/div, then center the signal on the display
with the delay knob (delay about 800
µ
s).
6
Press
Main/Delayed
, then press the
Delayed
softkey.
7
Set the delayed sweep to 20
µ
s/div, then set the expanded portion
over the VITS (delay about 988.8
µ
s).
Second VITS Displayed
This figure shows the second vertical interval test signals displayed with
delayed sweep.
Figure 6-7
Using Option 005 Enhanced TV/Video Trigger (HP 54645A)
To analyze video waveforms with Option 005
6-13