![HP 54645A Скачать руководство пользователя страница 229](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177229.webp)
To adjust the Option 005 offset (R15) (HP 54645A)
On the Option 005 board, there is only one adjustment. You must calibrate
the oscilloscope before performing this adjustment. You must also remove
the cabinet to access R15. Refer to chapter 8, “Replacement Parts,” for
instructions on how to remove the cabinet to gain access to R15.
Table 7-17
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Digital multimeter
0.1-mV resolution, accuracy
±
0.05%
HP 34401A
Adapter
BNC (m) to dual-banana post
HP 10110B
1
Set up the oscilloscope for the voltage adjustment.
a
Turn off the oscilloscope.
b
Remove the cabinet from the oscilloscope.
c
Connect the negative lead of the digital multimeter to a ground point
on the oscilloscope.
d
Connect the oscilloscope’s vertical output to the voltage inputs of the
digital multimeter (DMM) using the DMM’s test leads and the
BNC-to-dual-banana post adapter.
e
Turn on the oscilloscope.
f
Set up the digital multimeter for a DC voltage measurement.
g
Press
Setup
.
h
Press
Default Setup
.
2
Adjust R15 on the Option 005 PC board so that the measured voltage
on the digital multimeter is 0 volts
±
1 mV.
If adjusting R15 does not bring the voltage within tolerance, see “To
troubleshoot Option 005” later in this chapter.
Testing, Adjusting, and Troubleshooting
To adjust the Option 005 offset (R15) (HP 54645A)
7-40
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...