Testing Threshold Accuracy
Instructions for HP 54645D
These instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.
Testing the threshold accuracy verifies the data channel threshold
accuracy specification. You should perform this test every 24 months
or 4000 hours of operation, whichever comes first.
These instructions include detailed steps for testing the threshold
settings of digital channels D0-D7. After testing digital channels
D0-D7, connect and test digital channels D8-D15. Use the detailed
steps for digital channels D0-D7, substituting digital channels D8-D15
for digital channels D0-D7 in the instructions.
Each threshold test tells you to record the voltage reading in the
Performance Test Record located at the end of this chapter. To check
if each test passed, verify that the voltage reading you record is within
the limits listed on the Performance Test Record.
Table 7-2
Equipment Required
Equipment
Critical Specifications
Recommended Model/Part
Digital Multimeter
0.1 mV resolution, 0.005%
accuracy
HP 34401A
Function Generator
DC offset voltage
±
6.3 V
HP 3325B Option 002
BNC-Banana Cable
HP 11001-60001
BNC Tee
HP 1250-0781
BNC Cable
HP 10503A
BNC Test Connector, 8-by-2
7-12
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...