1
Set up the oscilloscope.
a
Press
Setup
, then press the
Default Setup
softkey.
b
Adjust the channel 1 Position knob to place the baseline at
approximately 0.5 division from the bottom of the display.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
c
Set the Volts/Div to the first line of the following table.
d
Press
Display
. Then press the
Average
softkey and set
# Average
to
64
. Wait a few seconds for the measurement to settle; the
Av
letters
in the status line indicate how much of the averaging process is
finished by turning to inverse video as the oscilloscope performs
averaging.
2
Press
Cursors
, then press the
V1
softkey.
3
Using the Entry knob, set the V1 cursor on the baseline of the signal.
4
Connect the power supply to the oscilloscope and to the multimeter,
using the BNC tee and cables.
5
Adjust the power supply output so that the multimeter reading
displays the first Power supply setting value in the following table.
Wait a few seconds for the measurement to settle.
6
Press the
V2
softkey, then position the V2 cursor to the center of the
trace.
The
∆
V value on the lower line of the display should be within the test limits
of the following table. If a result is not within the test limits, see
"Troubleshooting the Oscilloscope."
Testing, Adjusting, and Troubleshooting
To verify voltage measurement accuracy
7-17
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...