To verify vertical output on Option 005
This section applies only to the Option 005 Enhanced TV/Video Trigger.
In this test, we will use the oscilloscope’s channel 2 to measure the amplitude
of the vertical output signal.
Test limits: approximately 90 mV
p-p
into 50
Ω
with a full-screen input.
Table 7-10
Equipment Required
Equipment
Critical Specifications
Recommended Model/Part
Signal generator
1 to 500 MHz at 200 mV
(only 100 MHz needed)
HP 8656B opt 001
Cable
BNC, 48-inch
HP 10503A
Cable
Type N (m), 24-inch
HP 11500B
Adapter
Type N (m) to BNC (f)
HP 1251-0780
Feedthrough
50
Ω
, BNC (m) and (f)
HP 10100C
1
Connect the signal generator to the channel 1 input of the
oscilloscope.
2
Set the signal generator to equal the full bandwidth of your
oscilloscope, and set the output level to 0 dBm.
3
Connect the signal generator to the oscilloscope’s channel 1. Use a
50
−Ω
feedthrough to correctly terminate the signal generator.
4
Press
Autoscale
.
5
Adjust the oscilloscope controls and signal generator to obtain an
8-division-high display.
6
Connect the vertical output to channel 2. Use a 50
−Ω
feedthrough.
7
Measure the peak-to-peak amplitude of channel 2. It should be
greater than or equal to 63.6 mV
p-p
.
Because the measurement is being made at the full bandwidth of the
oscilloscope’s channel, the peak-to-peak measurement is corrected for the
oscilloscope’s high-frequency rolloff.
Testing, Adjusting, and Troubleshooting
To verify vertical output on Option 005
7-27
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...