To adjust the low-frequency compensation
In this procedure, you adjust the low-frequency compensation adjustment for
each channel.
Table 7-14
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Square wave
generator
30 kHz at about 3 V
p-p
HP 8112A
Feedthrough
50
Ω
, BNC (m) and (f)
HP 11048C
Cable
BNC
HP 10503A
1
Using the
BNC
cable and 50-
Ω
feedthrough, connect the square wave
generator to channel 1. Use the 50-
Ω
feedthrough at the oscilloscope
input.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
2
Adjust the generator for about 30 kHz at about 3 V
p-p
.
3
Press
Autoscale
.
4
Set channel 1 to 500 mV/div.
You must perform this adjustment on the 500-mV range.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
5
Adjust the output of the generator until you obtain about 5 to 6
divisions of vertical deflection.
Testing, Adjusting, and Troubleshooting
To adjust the low-frequency compensation
7-35
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...