HP 54645A/D Oscilloscope Digitizing System
Vertical Resolution
8 bits on all settings except 7 bits on < 2 mV/div to
1 mV/div
Horizontal Resolution
500 (displayed points)
Sample Rate
200 MSa/s maximum per channel
Sample Rate Accuracy
0.01%
Peak Detection
5 ns digital
Averages
4, 8, 16, 32, 64, 128, 256, Smoothing—selectable
Memory Depth
1 M maximum
Data Throughput
Up to 3.0 million samples per second with sufficient
trigger rate, and not using averaging, time base
vernier, waveform math, or vectors.
HP 54645D Logic Digitizing System
Vertical Resolution
1 bit
Maximum Horizontal Resolution
500 (displayed points)
Sample Period
2.5 ns maximum
Glitch Detection
8 channels
16 channels
5 ns
2.5 ns
5 ns
Memory Depth per Channel
2 M maximum
Sample Period Accuracy
0.01%
Simultaneous Capture
≤
8 channels on same pod
> 8 channels
any 2 channels on 2 pods
Available on all channels.
400 MSa/s
200 MSa/s
200 MSa/s
Acquire
All channels off
0-7 on, 8-15 off
8-15 on, 0-7 off
All on
channels 0-7 @ 400 MSa
channels 0-7 @ 400 MSa
channels 8-15 @ 400 MSa
channels 0-15 @ 200 MSa
Data Throughput
Up to 1.5 million samples per second with sufficient
trigger rate, and not using time base vernier.
Performance Characteristics
HP 54645A/D Oscilloscope Digitizing System
9-5
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...