![HP 54645A Скачать руководство пользователя страница 265](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177265.webp)
HP 54645A/D Time Base
Range
2 ns/div to 50 s/div (HP 54645A)
5 ns/div to 50 s/div (HP 54645D)
Resolution
40 ps
Vernier
1-2-5 increments when Off, 25 minor increments
between major settings when On
Reference Positions
Left, Center, Right
Random Repetitive
Stop freezes display
Pan and Zoom (Random Repetitive)
Shows only 1 trigger
Delay Range
Pre-trigger (negative delay)
Post-trigger (positive delay)
1 screen or 2.5 ms (the greater running repetitively)
500 seconds
Scope delta t accuracy
Same channel
Channel-to-Channel
(non-vernier ranges)
±
0.01% reading
±
0.2% screen width
±
40 ps
±
0.01% reading
±
0.2% screen width
±
80 ps
Logic delta t accuracy
Same channel
Channel-to-channel
(non-vernier ranges)
±
0.01% reading
±
0.2% screen width
±
(1 logic
sample period, 2.5 or 5 ns)
±
chan-to-chan skew
±
0.01% reading
±
0.2% screen width
±
(1 logic
sample period, 2.5 or 5 ns)
±
chan-to-chan skew
Delay Jitter
10 ppm
RMS Jitter
0.025% screen width + 100 ps pp
Entrance to Delayed Sweep
Not allowed when using any analog channels and
> 8 digital channels.
Performance Characteristics
HP 54645A/D Time Base
9-6
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...