![HP 54645A Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/hp/54645a/54645a_users-and-service-manual_166177052.webp)
To operate the time base controls
The following exercise guides you through the time base keys, knobs, and
status line.
•
Turn the Time/Div knob and notice the change it makes to the status
line.
The Time/Div knob changes the sweep speed from 2 ns to 50 s (HP 54645A)
and from 5 ns to 50 s (HP 54645D) in a 1-2-5 step sequence, and the value is
displayed in the status line.
•
You can access the horizontal
Vernier
softkey by pressing the
Main/Delayed
key.
The
Vernier
softkey allows you to change the sweep speed in smaller
increments with the Time/Div knob. These smaller increments are
calibrated, which result in accurate measurements, even with the
vernier turned on.
•
You can check the sample rate by pressing the
Main/Delayed
key
when the instrument is running, during a single run, or when the
instrument is stopped. The sample rate is also displayed when the
acquisition stops or ends.
•
Turn the Delay knob and notice that its value is displayed in the status line.
The Delay knob moves the main sweep horizontally, and it pauses at
0.00 s, mimicking a mechanical detent. At the top and bottom of the
graticule is a solid triangle (
▼
) symbol and an open triangle (
∇
)
symbol. The
▼
symbol indicates the trigger point and it moves with
the Delay knob. The
∇
symbol indicates the time reference point. If
the time reference softkey is set to left, the
∇
is located one graticule
in from the left side of the display. If the time reference softkey is set
to center, the
∇
is located at the center of the display. The delay
number tells you how far the reference point
∇
is located from the
trigger point
▼
.
Front-Panel Overview
To operate the time base controls
2-27
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...