9
Is the fan running?
If yes, go to "To run the internal self-tests." If not, do the steps below.
The Low Voltage Power Supply has a thermal cut-out circuit. If the fan is
defective, the Low Voltage Power Supply shuts down when it gets too hot for
safe operation.
a
Disconnect the fan cable from the power supply.
b
Measure the fan voltage at the connector on the power supply.
See the following figure for location of the fan connector. If the fan
voltage is approxi8 Vdc, replace the fan. If not, replace the
power supply.
Fan Connector
+5.25 V
adjust
-5.25 V
adjust
fan connector
Figure 7-10
Testing, Adjusting, and Troubleshooting
To check out the oscilloscope
7-45
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...