To capture glitches or narrow pulses with peak detect
A glitch is a rapid change in the waveform that is usually narrow as compared
to the waveform.
Peak detect mode displays any signal wider than 5 ns, allowing you to find
glitches and narrow spikes regardless of the sweep speed.
To capture glitches or narrow pulses:
1
Connect a signal to the oscilloscope and obtain a stable display.
2
Find the glitch. To select peak detect, press
Display
. Then,
press the
Peak Det
softkey.
Peak detect operates at sweep speeds from 50 s/div to 2 ns/div. When
operating, the initials Pk are displayed in the status line in inverse
video. At sweep speeds faster than 2 ns/div, the Pk initials are not
displayed in inverse video, which indicates that peak detect is not
operating.
Use Autostore to:
•
View waveforms that are changing.
•
View and compare waveforms with stored waveforms.
•
View narrow pulses or glitches that occur infrequently but require the
use of sweep speeds outside the range of peak detect.
•
Press
Autostore
.
Autostore retains the glitch on the display in half-bright video.
Making Measurements with HP 54645A/D Oscilloscopes
To capture glitches or narrow pulses with peak detect
5-11
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...