To verify bandwidth
In this test you verify bandwidth by using a power meter and power sensor to
set output of a signal generator at 1 MHz and at 100 MHz. You use the
peak-to-peak voltage at 1 MHz and at 100 MHz to verify the bandwidth
response of the oscilloscope. The test is repeated at 5 mV/div (75-MHz
bandwidth).
Test limits at 10 mV/div to 5 V/div:
all channels (
±
3 dB)
dc to 100 MHz
ac coupled 10 Hz to 100 MHz
Test limits at <10 mV/div:
all channels (
±
3 dB)
dc to 75 MHz
ac coupled 10 Hz to 75 MHz
Table 7-7
Equipment Required
Equipment
Critical specifications
Recommended Model/Part
Signal generator
1 to 100 MHz at 200 mV
HP 8656B opt 001
Power meter and
Power
Sensor
1 to 100 MHz
±
3% accuracy
HP 436A and HP 8482A
Power splitter
Outputs differ by < 0.15 dB
HP 11667B
Cable
Type N (m), 24 inch
HP 11500B
Adapter
Type N (m) to BNC (m)
HP 1251-0082
Feedthrough
50
Ω
, BNC (m) and (f)
HP 11048C
Testing, Adjusting, and Troubleshooting
To verify bandwidth
7-19
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...