HP 54645A/D Performance Test Record
Serial No. ______________________________________
Test by____________________________
Test Interval ____________________________________
Work Order No._____________________
Recommended Next Testing _______________________
Temperature _______________________
Threshold Accuracy
±
(100 mV + 3% of threshold setting)
Ch D0-D7
Ch D8-D15
Test
Specification
Limits
+5 V – 250 mV
+4.750 V
_____________
_____________
+5 V + 250 mV
+5.250 V
_____________
_____________
–5 V – 250 mV
–5.250 V
_____________
_____________
–5 V + 250 mV
–4.750 V
_____________
_____________
0 V – 100 mV
–100 mV
_____________
_____________
0 V + 100 mV
+100 mV
_____________
_____________
Calibrator Output
Nominal
Test Limits
Results
dc
0
µ
V
–500
µ
V to +500
µ
V
_____________
5 V
4.990 V to 5.010 V
_____________
Voltage measurement accuracy
Range
Power Supply Setting
Test Limits
Channel 1
Channel 2
5 V/Div
35 V
34.24 V to 35.76 V
_____________
_____________
2 V/Div
14 V
13.696 V to 14.304 V
_____________
_____________
1 V/Div
7 V
6.848 V to 7.152 V
_____________
_____________
500 mV/Div
3.5 V
3.424 V to 3.576 V
_____________
_____________
200 mV/Div
1.4 V
1.3696 V to 1.4304 V
_____________
_____________
100 mV/Div
700 mV
684.8 mV to 715.2 mV
_____________
_____________
50 mV/Div
350 mV
342.4 mV to 357.6 mV
_____________
_____________
20 mV/Div
140 mV
136.96 mV to 143.04 mV
_____________
_____________
10 mV/Div
70 mV
68.48 mV to 71.52 mV
_____________
_____________
5 mV/Div
35 mV
34.24 mV to 35.76 mV
_____________
_____________
2 mV/Div
14 mV
13.696 mV to 14.304 mV
_____________
_____________
1 mV/Div
7 mV
6.696 mV to 7.304 mV
_____________
_____________
Bandwidth
Test Limits
Channel 1
Channel 2
≤ ±
3 dB at 100 MHz
_____________
_____________
≤ ±
3 dB at 75 MHz
_____________
_____________
Horizontal
∆
t and 1/
∆
t accuracy
Generator Setting
Test Limits
Results
Frequency
10 kHz
9.96 kHz to 10.04 kHz
_____________
Period
100
µ
s
99.59
µ
s to 100.41
µ
s
_____________
Frequency
1 MHz
995.9 kHz to 1.0042 MHz
_____________
Period
1
µ
s
995.9 ns to 1.0041
µ
s
_____________
Frequency
100 MHz
98.61 MHz to 101.43 MHz
_____________
Period
10 ns
9.86 ns to 10.14 ns
_____________
Trigger sensitivity
Test Limits
Channel 1
Channel 2
Internal trigger
25 MHz at 0.35 divisions
_____________
_____________
100 MHz at 1 division
_____________
_____________
External
External trigger
100 MHz at 100 mV
p-p
_____________
25 MHz at 50 mV
p-p
_____________
Option 005 Voltage measurement accuracy
Test Limits
Channel 1
Channel 2
Amplitude peak-to-peak
≥
63.6 mV
p-p
_____________
_____________
7-28
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...