1
Connect the pulse generator to channel 1, using a 50-
Ω
feedthrough
at the oscilloscope input.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
2
Press
Autoscale
.
3
Change the time base to 10 ns/div.
4
Press
1
on the HP 54645A, or
A1
on the HP 54645D, then
toggle the
Vernier
softkey to
On
.
5
Adjust the Volts/Div until there are about 6 divisions of vertical
deflection.
6
Adjust the channel 1 high-frequency response for 1.0 minor division
of overshoot (4%).
7
Repeat steps 1 through 6 for channel 2.
Channel 2 is labeled "2" on HP 54645A and "A2" on HP 54645D.
High-Frequency Pulse Response Adjustment Locations
High-frequency
pulse response
adjustments
Figure 7-7
Testing, Adjusting, and Troubleshooting
To adjust the high-frequency pulse response
7-37
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...