Internal Trig
Sensitivity
1
Press
Setup
, then press
Default Setup.
2
Connect the signal generator to channel 1 using a 50-
Ω
feedthrough
at the oscilloscope input.
Channel 1 is labeled "1" on HP 54645A and "A1" on HP 54645D.
3
Verify the trigger sensitivity at 25 MHz and 0.35 divisions.
a
Set the signal generator to 25 MHz and about 100 mV.
b
Press
Autoscale
.
c
Set the time base to 50 ns/div.
d
Set channel 1 to 100 mV/div.
e
Decrease the output of the signal generator until there is 0.35 vertical
divisions of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope."
f
Record the result on the Performance Test Record as Pass or Fail.
4
Verify the trigger sensitivity at 100 MHz and 1 division.
a
Change the output of the signal generator to 100 MHz and set
amplitude to about 100 mV.
b
Set the time base to 10 ns/div.
c
Decrease the output of the signal generator until there is 1 vertical
division of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope."
d
Record the result on the Performance Test Record as Pass or Fail.
5
Repeat steps 1 through 4, substituting channel 2 for channel 1 in the
procedure.
Channel 2 is labeled "2" on HP 54645A and "A2" on HP 54645D.
Testing, Adjusting, and Troubleshooting
To verify trigger sensitivity
7-25
Содержание 54645A
Страница 5: ...iv ...
Страница 12: ...1 Getting Started ...
Страница 17: ...Items Supplied with the HP 54645A D Oscilloscopes Figure 1 3 Getting Started Check package contents 1 6 ...
Страница 18: ...Optional Accessories for the HP 54645A D Oscilloscopes Figure 1 4 Getting Started Check package contents 1 7 ...
Страница 25: ...1 14 ...
Страница 26: ...2 Front Panel Overview ...
Страница 60: ...3 Triggering HP 54645A D Oscilloscopes ...
Страница 84: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 142: ...5 Making Measurements with HP 54645A D Oscilloscopes ...
Страница 176: ...6 Using Option 005 Enhanced TV Video Trigger HP 54645A ...
Страница 189: ...6 14 ...
Страница 190: ...7 Testing Adjusting and Troubleshooting ...
Страница 242: ...8 Replacement Parts ...
Страница 254: ...Exploded View of HP 54645A D Oscilloscope Figure 8 3 Replacement Parts To order a replacement part 8 13 ...
Страница 259: ...8 18 ...
Страница 260: ...9 Performance Characteristics ...
Страница 271: ...9 12 ...
Страница 272: ...10 Messages ...