A-12
Histogram Overflow in datahist
Histogram Smoothing Error
Illegal Context Record Type
Illegal high value in qgen
Inconsistent d0 limits in rsattencalc
Indeterminate or bad data for attenuation
curves
Indeterminate Data For k Calculation
Insufficient Data To Determine k/delta0
Software
These messages should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Internal Buffer Size Exceeded
Software
This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Internal device request queue full [E_QUEUE_FULL_I]
Comment: The common buffer area for storing system sounds is full.
Action: Run CHEKPART to verify system software is correctly installed. Reduce the
size of one or more of the tunes or tune files you used to specify the system sounds.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS
COMMENT/ACTION