Discovery QDR Series Technical Manual
3-50
The correct display should show a flat graph with an amplitude of approximately 5.5 volts
with the alignment fixture installed (approximately 6.5 volts with alignment fixture
removed. volts. If thebeam alignment is not correct, perform the following procedure.
Figure 3-24
12. Set the machine in the Center Table position.