Bad checksum in message. [E_BAD_BLOCK_I]
Comment: The checksum calculated by the C/C for a message did not match the one in
the message packet.
Action: Either the driver is incorrectly configured, one or more devices are jumpered to
the debug mode, there is a hardware problem on the communication link (cable or
connectors) or one of more of the DSP chips in the C/C or DAS board is not properly
seated. Verify that the DEVICE=C:\MENU\MAPIDEV.SYS line in CONFIG.SYS has
the options /OVERLAP=0 and /HANDSHAKE=1. Check each stepper motor board
and verify that its address is correctly set. Verify that the TZ motor board does not
have the debug jumper installed. Verify that the communications cable is installed and
properly secured on the back of the C/C board. If the problem persists, it may be
because one (or both) of the DSP chips on the C/C or DAS board is not properly seated
in the PGA socket. Try "massaging" the DSP chip on the C/C board and the one on the
DAS board. If the problem changes (gets worse or better) when you do so, replace the
appropriate board. If the problem persists, place all devices in emulation mode (e.g., by
turning off instrument power and booting in the workstation mode). Disconnect all
devices except the C-arm from the communications link. Turn off C-arm emulation. If
there are no problems, proceed to re-connect each device and take it out of emulation
until the offending device is found. Replace the appropriate board.
BoundaryLine Internal Software Error
Software
This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Cannot generate bad detector map, too many bad detectors
Action: One or more detectors is bad. See instructions for running hardware checking
program.
Cannot obtain results of scan ________ of __________
Software
Comment: For some reason, the system was unable to obtain the results of a scan that
was selected for normals plotting.
Action: This message should be reported to Hologic Technical Support.
Can't open a window
Software
Action: Ensure that only Hologic software is installed on the computer. If so, report
this message to Hologic Technical Support. Include as much information as possible
about what you were doing at the time.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS
COMMENT/ACTION
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......