Discovery QDR Series Technical Manual
3-41
The program switches to graphics mode and draws the Encoder Vs Distance cal-
ibration grid. It steps the TX motor in by 1" increments until the inner mechani-
cal stop is hit. The program plots the encoder data during the scan (in raw
encoder units, 0-4095).
Press
<Esc>
at anytime during the scan to terminate the calibration procedure.
When the calibration completes, the program computes the linear fit for both the
positive and negative motion. The line for parameters are displayed at the top left
and top right of the plot in the form Y = Int Slope * X. The two slopes
(87.0 and 86.9 in the example below) should be within 0.3 of each other.
The program displays the positive and negative limits as horizontal dashed lines.
Note: The PosLimitOffset and NegLimitOffset entries in the [TxMotor] section
of the SQDRIVER.INI file determine the motion limits relative to the
mechanical stops. If these entries are not present, or are zero, the motion
limits are set to the mechanical limits.
7. The program then changes the plot title to PRESS ANY KEY TO CONTINUE.
Press the
<Enter>
key and the program prompts:
motor_direction=1
calibrate_position=1,2287,27446,43,12563,1026,1026,3096
pos_limit_position=604777
neg_limit_position=0
Update Driver INI-File [Y/N] ?
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......