Discovery QDR Series Technical Manual
9-5
9.1.3.2.5 0/1/2/3 buttons.
The default for these buttons is 0. The Field Engineer will not use the 0/1/2/3 buttons and
they should always remain the default.
9.1.3.2.6 0/1 buttons.
The default for these buttons is 0. The Field Engineer will not use the 0/1 buttons and they
should always remain the default.
Label Information
This area also provides information on the Hi Voltage Setting (either ON or OFF),
Frequency, Filter Errors and HiLo Errors.
9.1.3.3 Controls starting with Test Signals
Figure 9-6. Controls starting with Test Signals
These controls include:
9.1.3.3.1 Test Signal button OFF/ON.
The default setting of this button is OFF. When this button is ON a test signal is sent to all
of the detectors. Activity will be shown on the main display (see Figure 9-3 or Figure 9-4).
9.1.3.3.2 Pulse 0/1/2/3.
The default setting of this button is 1. This indicates what data is displayed and not what
the scanner is doing. The Field Engineer will not use the Pulse 0/1/2/3 buttons and they
should always remain the default.
9.1.3.3.3 Drum Half 0/1.
The default setting for this button is 0. This indicates what data is displayed and not what
the scanner is doing. The Field Engineer will not use the Pulse 0/1 buttons and they should
always remain the default.
AW-00335_001-0906
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......