5-1
Section 5
REMOVE AND REPLACE
PROCEDURES
This section describes how to remove and replace the Field Replaceable Units (FRUs) in the
Discovery. To safely perform a FRU removal or replacement, take care to follow the procedure
precisely as written.
Note: Whenever a component in the x-ray path is replaced, you must recalibrate and rerun QC.
5.1 Recommended Tools
5.2 Electronics Tray FRUS
This section describes how to remove and replace the FRUs in the Electronics Tray/Carriage
Drive area of the Discovery (see Figure 5-1). To remove any of the FRUs in the Electronics Tray
assembly, remove the 6 Phillips retaining screws and remove the cover. Next, remove the nut that
holds the ground wire to the cover ground lug.
Tool
Size/Type
Hex driver
3/32"
Hex driver or wrench
5/32"
Nut driver
1/4"
Nut driver
3/8 inch
Nut driver
5/16"
Nut driver
7/16"
Screwdriver
Narrow slotted
Screwdriver
Phillips head
Screwdriver
Slotted
Wrench
3/8 inch
ESD PRECAUTIONS: To prevent damage due to ESD (Electrostatic Discharge), you
must take precautions when handling components. Remove any charge from your body
by wearing an approved and properly grounded wrist strap. Keep PCBs in their ESD
protective bag until you are ready to install them. Treat defective PCBs as new to
prevent any additional damage.
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......