Discovery QDR Series Technical Manual
3-7
Figure 3-6. Crated Unit (Discovery C, Ci, W and Wi)
3.2.2 Uncrate Unit
Remove the unit from the crate as described below:
1. Cut the strapping that holds the packaging together.
2. Remove the cardboard cap by lifting it up and off.
3. Remove the cardboard sleeve by lifting it straight up and off. Be careful not to
scratch the unit.
4. Remove the wooden table shelf and the boxes packed with the unit.
Note: There is a metal cross brace attached to the wooden table shelf. Remove
and save this cross brace as it is used later in the installation.
5. Verify that the serial numbers on both crates and the scanner all match.
3.2.2.1 Inspect For Hidden Shipping Damage
Open all crates and boxes, and check for signs of hidden damage. Check the ShockWatch
and Tip-N’-Tell indicators for evidence of improper handling during shipment. Bring any
damage discovered to the attention of the customer's Shipping/Receiving department.