Discovery QDR Series Technical Manual
5-35
Figure 5-18. Rear Drum Assembly FRIS
1. Shut down the computer and then turn off the main circuit breaker on the foot end
pedestal.
2. Remove the Aperture Assembly by removing 4 Phillips screws that secure the
assembly, and 2 Phillips screws that secure the aperture motor mount, to the drum
end plates.
3. Unplug the PCB cable.
4. Remove and replace the PCB (ensure that the encoder wheels are not touching
the sensors.
5. Attach Ch1 probe to SEGMENT test point (TP3) on the C-Arm Interface board.
6. Attach Ch2 probe to BRASS test point (TP2) on the C-Arm Interface board.
7. Connect probe to ground.
8. Go to X-Ray Survey and turn on the Filter Drum motor.
9. As the filter drum motor turns on and rotates look on the C-Arm Interface board.
a. The yellow “Index” LED blinks (on then off) as the Index mark passes the
optical interrupter.
b. Verify the green “Top of Drum” and “Brass on Top” LED (D8) blink as the
drum spins.
c. The green “AC Lock” LED (D7) lights (steady, not blinking) after a few rev-
olutions of the drum.
10. Measure from the rising edge of the BRASS to the rising edge of the SEGMENT
(it does not matter which edge leads). The time measured must be less than 500
μ
s (500
μ
s = 0.5 ms). See Figure 5-19 and Figure 5-20.
Encoder Wheel, Inner
Encoder Wheel, Outer
Drum Encoder PCB
140-0065
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......