Discovery QDR Series Technical Manual
7-2
[ ] On A and W system models, perform the Table Top Radiographic Uniformity test.
Print a copy of the test results and include it with the other paperwork being returned
to Hologic.
[ ] Run 10 spine scans and check calibration (do not recalibrate unless necessary).
[ ] Print copies of the baseline phantom scan, recent phantom scan, and the results of scan
averaging for the 10 spine scans in the previous step. Compare these printouts with
those from the last PM for possible problems, and send the printouts to Hologic to be
kept with the service records for this system.
[ ] From the Discovery Main Menu, select
QC|QC Data Management|Plot
. Setup the
QC Plot to include QC scans from at least the last wo years. Display and print a copy
of the BMD, BMC, Area, HiAir, and LoAir plots to be included in the paperwork
being returned to Hologic.
[ ] During a scan, verify that pressing the red Emergency STOP switch immediately stops
all C-Arm motion, table motion, and X-ray production. X-ray production should be
monitored by a Victoreen 450P or equivalent.
[ ] Test all tape switches. The tape switches are located along the length of the front and
back of the table, and mounted on the frame under the table. Verify that pressing a tape
switch immediately stops all C-Arm motion, table motion, and X-ray production. X-
ray production should be monitored by a Victoreen 450P or equivalent.
Note: The front (table) tape switch is disabled until the C-arm is at 0
°
, so this switch
should be checked with the C-arm at a position greater than 30
°
.
[ ] Replace (if necessary) and adjust the motor drive belts as follows:
Note: Detailed instructions for drive belt replacement and adjustments can be found
in the REMOVE AND REPLACE section of this manual.
[ ] For each drive belt (C-Arm Y, C-Arm R, Table X, and Table Y) perform the following:
•
Loosen the 2 mounting bolts holding the tension block.
•
Tighten the tension nut so that the spring is compressed to 7/8 inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut). For the AR tension spring, adjust to 7/8" from the inside of
one washer to the inside of the other.
•
Tighten the two mounting bolts holding each drive belt’s tension block.
Note: It is not necessary to perform the MOTOR calibration procedures after
tensioning the motor drive belts.
[ ] Ensure that all cable connections are tight.
[ ] Run CHECKDISK on the C drive. This utility program can be started by clicking on
start|Run…
, then typing
CHKDSK
, and then clicking the
OK
button. DO NOT run
Scandisk on any Archive media.
[ ] Run Disk Defragmenter to and compact and re-order the files on the computer's hard
disk. This utility program can be started by clicking on
Start|Run…
, then typing
DEFRAG
, and then clicking the
OK
button.
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......