X-ray signal level (hi/lo) did not match the filter drum position [E_XRAY_PHASE_I]
Comment: If the filter drum monitor is enabled and the filter drum is on, this error
indicates that the filter drum wheel readback did not match the expected value (i.e.,
there must be six segment readbacks, there must be eight pulses within each segment
and the first four pulses in each segment are required to be brass and the second four
are required to be air).
Action: See the discussion under E_XRAY_SEQUENCE_I.