Discovery QDR Series Technical Manual
4-38
4.14.4 RECALYZE and Add Array AP Scans to the QC
Database
Once the unit has been calibrated, AP scans must be re-analyzed and added to the QC
database.
After the machine has been calibrated for thickness measurement in the array mode, it
must be calibrated for the array AREA and BMC measurements.
1. From the Discovery Main Menu screen, select
Utilities|Service Utili-
ties|Auto Analyze
.
2. Select Analyzed scans, click
RECALYZE
, and select the 20 scans just acquired
and click the
Next
button.
3. . The scans will now be recalculated to the new calibration.
4. Highlight the scans from the Excluded list and click the
Include Scans
button
and then the
Finish
button. (Selection is easier if you click Scan Date bar above
the list. This will order the scans by date and time).From the Discovery Main
Menu screen, select
QC|QC Data Management|Select Scans
.
5. From the Discovery Main Menu, select the
QC|QC Data Management|Plot
to create plots for BMD, BMC, AREA, HiAir, and LoAir plot for each array
mode. Verify that the QC plots inlcude QC scans from at least the last two years.
6. Ensure that all scans fall between the two, dotted limit lines.
7. Print the BMD, BMC, Area, HiAir, and LoAir plots. Include them with the other
paperwork being returned to Hologic.
Note: For more details on performing the QC setup and producing QC plots, refer to the
Discovery User's Guide.
Sample HARDWARE.INI (Calibration) File
HARDWARE.INI Variables
Set by…
Descriptions
[Version]
.INI File Section Name
HardwareModel=14
Factory
Discovery model (C=20, W=21, A=23, SL=22).
[Laser]
LaserYOffset=-38100
Factory
Laser Offset Value
[TxMotor]
PosLimitPosition=409575
Factory
Positive limit in microns (= 16 1/8 inches)
SpineInitPos=208000
Factory
TX center position in microns (= 8.18 inches)
MaxSpinePos=233400
Factory
+ Limit for spine placement
MinSpinePos=182600
Factory
- Limit for spine placement
PatientUnloadPos=0
Factory
Table position for unloading patients
APLatDiff=376200
Factory
AP/Lateral Difference
MaxTrackOffset=25400
Factory
Содержание DISCOVERY A
Страница 52: ...Discovery QDR Series Technical Manual 2 30...
Страница 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Страница 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Страница 144: ...Discovery QDR Series Technical Manual 4 22...
Страница 188: ...Discovery QDR Series Technical Manual 5 22...
Страница 228: ...Discovery QDR Series Technical Manual 6 16...
Страница 248: ......