![Hioki IM7580 Скачать руководство пользователя страница 201](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306201.webp)
193
Common Functions (LCR Mode, ANALYZER Mode)
7.3.7
Initializing the Instrument
Initialization of the setting is performed as follows.
Check “Instrument malfunction” (p. 297) if the instrument malfunctions.
Perform a system reset to restore the instrument to its factory default settings if the cause is not
known.
See “Appx. 8 Initial Settings Table” (p. A15).
A system reset can also be performed with the
*RST
,
:PRESET
,
:SYStem:RESet
communication
command.
See the included LCR Application Disk. - Communication Commands “
*RST
”, “
:PRESET
”,
“
:SYStem:RESet
”
CAUTION
Performing a system reset causes the instrument to return to its default factory settings.
Disconnect the measurement sample before performing a system reset.
Perform a full reset if the initialization screen cannot be displayed. (
1
2
1
Press
[SETUP]
.
2
Press
[COMMON]
tab.
Press
[ADVANCED]
tab for the ANALYZER mode.
7
Application Function
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......