132
Equivalent Circuit Analysis Function
4.9.3
Performing Equivalent Circuit Analysis
(1) Performs frequency sweep measurement
Before performing equivalent circuit analysis
Set the sweep parameter to “frequency” and acquire frequency characteristics of elements to be
analyzed.
Refer to “4.2.5 Setting the Sweep Parameter” (p. 74)
Because the local maximum and local minimum measurement points are used when performing
equivalent circuit analysis with this instrument, the frequency range should be set to the range
for which the local extreme values can be measured. Since low frequency values are used when
performing analysis with Model B or Model C, configure the settings so that the lowest possible
frequencies are measured.
Additionally, when performing analysis using E model, set the range so that it includes the
resonance points for series resonance and parallel resonance.
Examples of appropriate sweep range settings
Examples of inappropriate sweep range settings
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......