Sweep Measurement
Setting example for normal sweep and segment sweep
Sweep setting
items
Normal sweep
Segment sweep
Segment
Segment 1
Segment 2
Segment 3
Sweep parameter
Frequency
Frequency
Frequency
Frequency
Sweep range
1.0000 MHz to
300.00 MHz
1.0000 MHz to
5.0000 MHz
10.000 MHz to
50.000 MHz
50.000 MHz to
300.00 MHz
Number of sweep
points
801 points
201 points
201 points
399 points
Setting method
for sweep points
Log
Log
Log
Linear
Measurement
signal type
POWER
POWER
POWER
POWER
Measurement
signal level
0.0 dBm
0.0 dBm
1.0 dBm
-1.0 dBm
Average
5 times
10 times
3 times
OFF
Measurement
speed
FAST
FAST
MEDIUM
SLOW
Point delay
0.0005 s
0.0005 s
0.0010 s
0.0000 s
Normal sweep
Normal sweep range
Segment sweep
Segment 1
Segment 2
Segment 3
4
Analyzer Function
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......