72
Setting Basic Settings of Measurement
4.2.4
Applying the Signal to the Sample during Measurement Only
(Trigger Synchronous Output)
This function enables the measurement signal to be output for only the initial sweep point after
measurement is triggered, so that the signal is applied to the sample during measurement only. You
can also set a delay time to ensure that data is acquired after the sample stabilizes.
Thus reducing the generation of heat in the sample and decreasing electrode wear.
Trigger synchronous output: OFF
Trigger synchronous output: ON
Measurement
Calculation
TRIG
―
―
―
―
EOM
―――
INDEX
―
―――
―
Signal
Measurement
Calculation
Synchronous
signal wait
INDEX delay
TRIG
―
―
―
―
EOM
―――
INDEX
―
―――
―
Signal
1
3
2
1
Press
[SETUP]
.
2
Press
[SWEEP]
tab.
3
Press
[SYNC]
.
4
5
4
Press
[SYNC]
.
5
Select ON/OFF for the trigger synchronous output.
[OFF]
Disables the trigger synchronous output.
[ON]
Enables the trigger synchronous output.
Go to the next page.
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......