General Specifications
(2) LCR function
Measurement with single condition
Average
1. Method
Internal trigger: Moving average
External trigger: Arithmetic mean
2. Setting range 1 to 256 (1 step)
Trigger
1. Internal
trigger
Automatic
2. External
trigger
Manual, communication commands, I/O
Trigger delay
Delay time from trigger to measurement
0.00000 s to 9.99999 s (resolution: 10 µs)
Trigger
synchronous
output
Applies measurement signal during analog measurement only.
Stabilizing wait time setting
0.00000 s to 9.99999 s (resolution: 10 µs)
INDEX signal delay time
setting
0.00000 s to 0.10000 s (resolution: 10 µs)
BIN judgment
10 classifications for 4 items, OUT OF BINS
EXT I/O output available
1.
Upper lower limit value judgment
Upper lower limit
setting range
-9.99999 G to +9.99999 G
2.
Percent (%) judgment
Reference setting
range
-9.99999 G to +9.99999 G
Upper lower limit
setting range
-999.999% to +999.999%
3.
Deviation percentage (∆%) judgment
Measurement values are indicated as deviations (∆%) from the reference values.
Reference setting
range
-9.99999 G to +9.99999 G
Upper lower limit
setting range
-999.999% to +999.999%
Comparator
Hi/IN/Lo for 4 items
EXT I/O output available
1.
Upper lower limit value judgment
Upper lower limit
setting range
-9.99999 G to +9.99999 G
2.
Percent (%) judgment
Reference setting
range
-9.99999 G to +9.99999 G
Upper lower limit
setting range
-999.999% to +999.999%
3.
Deviation percentage (∆%) judgment
Measurement values are indicated as deviations (∆%) from the reference values.
Reference setting
range
-9.99999 G to +9.99999 G
Upper lower limit
setting range
-999.999% to +999.999%
Magnification
display function
The display of measurement values and comparator judgment results can be magnified.
12
Specifications
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......