![Hioki IM7580 Скачать руководство пользователя страница 205](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306205.webp)
197
External Input/Output Connector and Signals
Pin
I/O
Signal name
Function
Logic
Common
COMP
BIN
14
OUT
AND
―――
•
Outputs results obtained by applying an
AND operation to the judgment results for
measurement results of the four parameters.
• Output if all the judgment results is IN (parameters
not for considered for judgment are excluded).
Neg Level
BIN7
―
―
―
―
Output if the BIN judgment result is BIN7.
15
OUT
PARA3-IN
―
―――
―
―
―
―
Output if the third parameter comparator judgment
result is IN.
Neg Level
BIN9
―
―
―
―
Output if the BIN judgment result is BIN9.
16
OUT
PARA4-HI
―
―――
―
―
―
―
Output if the fourth parameter comparator
judgment result is HI.
Neg Level
17
OUT
PARA4-LO
―
―――
―
―
―
―
Output if the fourth parameter comparator
judgment result is LO.
Neg Level
18
OUT (Unused)
-
-
-
19
OUT
OUT_OF_BINS
――
―
―
――
―
―
―
――
BIN judgment results
Neg Level
20
IN
(Unused)
-
-
-
21
IN
(Unused)
-
-
-
22
IN
LD0
―
―
―
Neg Level
23
IN
LD2
―
―
―
Select panel No. (p. 203)
Neg Level
24
IN
LD4
―
―
―
Select panel No. (p. 203)
Neg Level
25
IN
LD6
―
―
―
Select panel No. (p. 203)
Neg Level
26
IN
LD_VALID
―
―
―
―
―
―
―
―
Neg Level
27
-
ISO_COM
―
――
―
―――
Isolated power supply common
-
-
28
OUT
EOM
―――
Measurement complete signal
(When this signal is output, the comparator
judgment results have been finalized.)
Neg Edge
29
OUT
INDEX
―
―――
―
•
Signal that indicates measurement (calculations
and judgment have not been processed) has
been completed.
•
The sample can be switched once this signal
changes from HIGH (OFF) to LOW (ON).
Neg Edge
30
OUT
PARA1-IN
―
―――
―
―
―
―
Output if the first parameter comparator judgment
result is IN.
Neg Level
BIN2
―
―
―
―
Output if the BIN judgment result is BIN2.
31
OUT
PARA2-HI
―
―――
―
―
―
―
Output if the second parameter comparator
judgment result is HI.
Neg Level
BIN4
―
―
―
―
Output if the BIN judgment result is BIN4.
32
OUT
PARA2-LO
―
―――
―
―
―
―
Output if the second parameter comparator
judgment result is LO.
Neg Level
BIN6
―
―
―
―
Output if the BIN judgment result is BIN6.
33
OUT
PARA3-HI
―
―――
―
―
―
―
Output if the third parameter comparator judgment
result is HI.
Neg Level
BIN8
―
―
―
―
Output if BIN judgment result is BIN8.
34
OUT
PARA3-LO
―
―――
―
―
―
―
Output if the third parameter comparator judgment
result is LO.
Neg Level
BIN10
―
―
―
―
―
Output if BIN judgment result is BIN10.
35
OUT
PARA4-IN
―
―――
―
―
―
―
Output if the fourth parameter comparator
judgment result is IN.
Neg Level
36
OUT (Unused)
-
Neg Level
37
OUT (Unused)
-
Neg Level
8
External Control
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......