278
General Specifications
(4) Continuous measurement function
Measurements are continued with measurement conditions saved.
1. Up to 46 ways
LCR mode: Up to 30 ways, ANALYZER mode: Up to 16 ways
Continuous measurements with LCR mode mixed with ANALYZER mode is possible
2.
Judgment result from EXT.I/O has an overall judgment result and more than one output pattern.
(5) Function
Contact check
1.
2-terminal contact check (DCR measurement)
Performs a contact (contact state) check between High and Low.
Judgment is allowed by entering upper and lower limit for DCR values.
A function that aborts subsequent measurements when the judgment result is FAIL available.
Check timing can be changed.
BEFORE: Contact check performed before measurement
AFTER: Contact check performed after measurement
BOTH: Contact checks performed before and after measurement
Measurement
a. Range 0.1
Ω
to 100
Ω
b. ±1.6% rdg. (typical)
Measurement signal
Level: 1 mA or less
Number of waveforms: 1 to 9,999 (8 µs to 79992 µs)
Wait time
Wait before DC measurement: 0 s to 9.99999 s
Wait before AC measurement: 0 s to 9.99999 s
AC signal superimpose available
2.
Hi-Z reject function (detection of OPEN state during 2-terminal measurement)
When the measurement value is higher than the judgment reference, a contact error is
output.
Judgment standard: Setting 1
Ω
to 10 k
Ω
(resolution: 1
Ω
) available
Error output: Error output from EXT I/O
3.
Waveform identification function (chattering detection)
Effective values of subsequent waveforms is compared with the effective value of the
waveform that is read first. A contact error is output if fluctuation of the subsequent waveform
exceeds the judgment reference.
Judgment reference: Setting 0.01% to 100.00% (0.01% resolution) with respect to the
reference value is possible.
Error output: Error is displayed on the LCD display and error is output from EXT I/O
Panel save
and load
function
Full measurement condition: Saving 30 types (LCR mode) and 16 types (ANALYZER mode) of
setting conditions are possible.
Compensation value only: Saving 30 types (LCR mode) of compensation values are possible.
Arbitrary measurement conditions can be read by key operations or a control signal via the EXT I/
O.
Display
digits setting
function
Number of display digits for measurement values can be set to 3, 4, 5, and 6. However, differs
based on the parameter. (default: 6 digits)
Display
Setting
function
1. LCD display ON/OFF (no drawing in case of OFF)
2.
Back light brightness adjustment
3. Measurement screen color customization (color with white background or black background)
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......