152
Calibration
5.2.2
Setting the Electric Length [LENGTH]
Compensation is performed for the error caused by the phase shift occurring between the
calibration reference surface and the measurement sample connection surface.
Enter the electric length between the calibration reference surface on which open calibration, short
calibration, and load calibration were performed for the surface where the measurement sample is
connected.
The guaranteed accuracy range differs depending on the electric length.
See “Electrical length compensation” (p. 279).
1
1
Press
[ADJUST]
.
2
2
Press
[LENGTH]
.
The numeric keypad can be
used for input.
4
3
3
Set the electric length with
/
or the numeric
keypad.
(when the numeric keypad is used, press
[SET]
.)
[C]
Repeats the input.
[CANCEL]
Cancels the setting.
4
Press
[EXIT]
to close the setting screen.
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......