117
Judging Measurement Results (Comparator Function)
4
A message such as the following may be displayed when
[MEAS VAL]
is selected.
In this case, set trigger settings to
[SEQ]
and perform a sweep once.
TRIG setting is REPEAT
Measured values cannot be referenced correctly because
trigger setting is REPEAT.
Some points have no Meas Value
Measured values cannot be referenced correctly because
there is a sweep point where the measured value is
invalid.
4
Sets the judgment area for sweep parameter 1.
[MEAS VAL]
Sets the upper and lower limit values with
the current measured values as reference.
[FIX VAL]
Sets the reference value, upper limit value,
and lower limit value.
[%]
Sets the upper and lower limit values as
percentage values relative to the reference
value.
[VAL]
Sets the upper and lower limit values as
absolute values relative to the reference value.
5
5
Displays the segment number to be set with
/
or by scrolling.
Only one row is displayed when the segment function
is OFF.
6
6
(This is enabled only when the setting of judgment
area is
[FIX VAL]
.)
(1)
Press the cell corresponding to REF of any
arbitrary segment.
(2)
Set the reference value with the numeric
pad* and press
[SET]
.
Settable range
-9.99999 G to 9.99999 G
7
7
(1) Press the cell corresponding to HI of any
arbitrary segment.
(2)
Set the upper limit value with the numeric
pad* and press
[SET]
.
Settable range
(set as % value)
-999.999% to 999.999%
Settable range
(set as absolute
value)
-9.99999 G to 9.99999 G
Go to the next page.
4
Analyzer Function
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......