144
Calibration
6
6
For each of
[OPEN]
,
[SHORT]
, and
[LOAD]
, the
offset delay values
*
of a random calibration kit is
set with the numeric keypad.
[RESET]
The set value becomes 0.
[CANCEL]
Cancels the setting.
Example: In case of
[OPEN]
7
6
7
Press
[SET]
.
[C]
The numerical value is entered again.
8
8
Press
[FORMAT]
.
9
10
9
Select the input parameter pattern for the
reference value set by
[OPEN]
,
[SHORT]
, and
[LOAD]
respectively.
See “3.2.1 Setting Display Parameters” (p. 32).
[OPEN]
G-Cp, G-B
[SHORT]
Rs-Ls, Rs-X
[LOAD]
Z-
θ
, Cs-D, Rs-Cs, Cp-D, Rp-Cp, Ls-Q,
Rs-Ls, Lp-Q, Rp-Lp, Rs-X
10
Press
[EXIT]
to close the setting screen.
What is an offset delay value?
The offset delay value is the one-way propagation time (s) from the calibration surface of the standard
unit to the definition surface. It affects the definition value. Use the value provided in the calibration kit
specifications.
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......