![Hioki IM7580 Скачать руководство пользователя страница 135](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306135.webp)
127
Equivalent Circuit Analysis Function
(3) Setting the frequency range for analyses
This section describes how to set the frequency range used to perform equivalent circuit analysis
when using normal sweep. This function allows you to restrict local extreme values to be used for
the analysis in case several local extreme values exist in the sweep range. Configures the setting
so that local extreme values are included in the analysis range. This setting is valid only during
normal sweep operation.
1
2
3
1
Press
[SETUP]
.
2
Press
[CIRCUIT]
tab.
3
Press
[AREA]
.
4
5
6
Changing the unit:
M
(mega)/
k
(kilo)
4
Press
[START]
, use the numeric keypad to enter
the frequency at which to start analysis, and press
[Hz]
.
Settable range 1.0000 MHz to 300.00 MHz
[C]
Repeats the input.
[CANCEL]
Closes the window without setting the
analysis range.
[RESET]
Resets the analysis range that has been
set.
5
Press
[STOP]
, use the numeric keypad to enter the
frequency at which to end analysis, and press
[Hz]
.
Settable range 1.0000 MHz to 300.00 MHz
6
Press
[SET]
to accept the frequency range.
The accuracy of the analysis may deteriorate if a very narrow frequency range is set.
4
Analyzer Function
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......