148
Calibration
Perform measurement
Performs the calibration measurement.
To avoid improper connection of the standard units, perform “Prevention of improper standard unit
connection” (p. 150) setting in advance.
Open measurement
2
1
Connects the standard unit for open to the test
sample connection terminal.
2
Press
[OPEN]
.
3
3
Press
[AC + DC]
.
Starts measurement.
If the standard unit used in AC measurement is
different from that the one used in DC measurement,
press
[AC]
,
[DC]
.
The results are displayed under
[OPEN]
after
completion of measurement.
[CANCEL]
Cancels the setting, and closes the screen.
Short measurement
2
1
Connect the standard unit for short to the test
sample connection terminal.
2
Press
[SHORT]
.
Содержание IM7580
Страница 1: ......
Страница 2: ......
Страница 6: ...Contents iv...
Страница 8: ......
Страница 18: ...10 Operating Precautions...
Страница 68: ...60 Judging Measurement Results...
Страница 230: ...222 External Control I O Settings...
Страница 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Страница 280: ...272 Editing Data Saved in USB Flash Drive...
Страница 300: ...292 Measurement Time...
Страница 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 328: ...A16 Initial Settings Table...
Страница 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 330: ...A18 Initial Settings Table...
Страница 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Страница 337: ...13 09...
Страница 338: ......
Страница 339: ......