Manufacturer’s declaration - Electromagnetic Compatibility
Electromagnetic emissions compliance
Electromagnetic immunity compliance
Manufacturer's recommendations and declaration – electromagnetic emissions
The Primo
™
is intended for use in the electromagnetic environment specified below. The customer or
the user of the Primo
™
should make sure it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - Recommendations
RF emissions
CISPR 11
Group 1
The Primo
™
only uses RF energy internally.
Consequently, it only produces very weak RF emissions
that are unlikely to cause interference with nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The Primo
™
is suitable for use in all establishments,
including domestic premises and premises directly
connected to the public low-voltage power supply in
residential buildings.
Harmonic emissions
IEC 61000-3-2
Class A
Emissions of voltage
fluctuations/flicker
IEC 61000-3-3
Compliant
Manufacturer's recommendations and declaration – electromagnetic immunity
The Primo
™
is intended for use in the electromagnetic environment specified below. The customer or
the user of the Primo
™
should make sure it is used in such an environment.
Immunity test
Test level
IEC 60601
Compliance
Electromagnetic environment -
Recommendations
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV on contact
±8 kV in the air
±6 kV on contact
±8 kV in the air
Floors should be made of wood,
concrete or ceramic tiles. If the floor is
covered with a synthetic material,
relative humidity must be at least
30%.
Fast electric
transients in
bursts
IEC 61000-4-4
±2 kV on power
supply lines
±1 kV for the
input/output lines
±2 kV on power
supply lines
±1 kV for the
input/output lines
The quality of the electric power
supply should be the equivalent of
that found in commercial or hospital
environments.
Shock waves
IEC 61000-4-5
±1 kV between line(s)
±2 kV between line(s)
and earth
±1 kV between line(s)
±2 kV between line(s)
and earth
The quality of the electric power
supply should be the equivalent of
that found in commercial or hospital
environments.
Magnetic field
at the
frequency of
the mains
power supply
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
The magnetic field at the frequency of
the mains supply must be of a level
that is characteristic of a place
located in a commercial or hospital
environment.
Содержание Primo P02033
Страница 2: ...170924 2 Primo PROPRIETARY AND CONFIDENTIAL DRAFT octobre 11 2012...
Страница 28: ...Page 22 170924 2 EN Primo User Manual PROPRIETARY AND CONFIDENTIAL DRAFT octobre 11 2012...
Страница 72: ...Page 22 170924 2 FR Manuel d utilisation Primo PROPRIETARY AND CONFIDENTIAL DRAFT octobre 11 2012...
Страница 116: ...Seite 22 170924 2 DE Primo Gebrauchsanweisung URHEBERRECHTLICHGESCH TZTERUNDVERTRAULICHERENTWURF...
Страница 160: ...Pagina 22 170924 2 NL Primo Gebruikershandleiding AUTEURSRECHTELIJKBESCHERMDENVERTROUWELIJKCONCEPT o...
Страница 204: ...Sida 22 170924 2 SV Anv ndarhandbok Primo PROPRIETARY AND CONFIDENTIAL DRAFT octobre 11 2012...
Страница 248: ...Side 22 170924 2 DA Primo Brugermanual OPHAVSRETLIGT BESKYTTET OG FORTROLIGT UDKAST octobre 11...
Страница 267: ...PROGETTO RISERVATO octobre 11 2012 IT 170924 Manuale d uso Primo P02033 P02034 P02044 P02047 170924 Rev 002...
Страница 292: ...Pagina 22 170924 2 IT Manuale d uso Primo PROGETTO RISERVATO octobre 11 2012...
Страница 336: ...P gina 22 170924 2 ES Manual de Usuario Primo DOCUMENTO PRIVADO Y CONFIDENCIAL octobre 11 2012...
Страница 380: ...P gina 22 170924 2 PT Manual do utilizador Primo PROJECTO PATENTEADO E CONFIDENCIAL octobre 11 2012...