GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
8-86
Section 8-20 - Replacement of the Voluson Station Components
8-20-25
Replacement of the VGA Connector
8-20-25-1
Manpower
One person, 10 minutes
8-20-25-2
Tools
Phillips screwdriver 1
8-20-25-3
Preparations
1.) Power Off/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
2.) Remove the complete lever.
3.) Remove the Top Shelf Bottom Rear Cover as described in
Section 8-20-13 on page 8-73
.
8-20-25-4
VGA Connector - Removal Procedure
1.) Screw off 1 screw.
2.) Disconnect the VGA cable.
8-20-25-5
VGA Connector - Installation Procedure
1.) Connect the VGA cable to the new VGA Connector and place it on its original position.
2.) Fasten it with the screw.
3.) Mount the Top Shelf Bottom Rear Cover as described in
Section 8-20-13-5 on page 8-73
.
4.) Mount the complete lever at its original position.
Figure 8-108 screw off the screw