Section 6-3 - LCD Display (Monitor) Adjustment
Section 6-3
LCD Display (Monitor) Adjustment
6-3-1
Color Calibration
1.) Press the
UTILITIES
key on the keyboard and then select the SYSTEM item.
2.) Select the
SERVICE
page, enter the password
SHE
and click the ACCEPT button.
3.) Select the MONITOR TEST button in the “Service Tools” menu.
4.) Confirm the “Monitor Test” message with YES.
The Monitor Test main menu appears on the screen.
5.) Select the “Color calibration” icon and press the
right/left trackball
(screen becomes white).
-
WHITE is displayed without any tint (discolor) or colored pixels.
6.) Press the
right/left trackball
key repeatedly to step through RED, GREEN and BLUE.
-
Each color is displayed correctly (without any tint or discolored pixels).
7.) To return to the Main menu, press the
upper trackball
key.
8.) To exit the Monitor Test program, press
EXIT
.
Figure 6-1 Monitor Test - Main menu
EXIT