Chapter 8 - Replacement Procedures
8-27
Section 8-9
Replacement of the Rotary Knob
8-9-1
Manpower
One person, 5 minutes
8-9-2
Tools
small-sized slotted screwdriver or tweezers
8-9-3
Preparations
1.) Power Off/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
8-9-4
Rotary Knob - Removal Procedure
1.) Carefully place a small flat screwdriver between the Rotary Knob and the surface of the keyboard
- see
Figure 8-32
below.
2.) Gently lift the defective Rotary Knob, until it is completely loosened from its base.
3.) Remove the defective knob.
8-9-5
Rotary Knob - Installation Procedure
1.) Place a new Rotary Knob in position on the keyboard, taking care to place the plastic alignment pin
in the correct location.
2.) Push the Rotary Knob down until it snaps into position.
Figure 8-32 Rotary Digipot Replacement