Chapter 3 - Setup Instructions
3-29
3-5-2
Connection and Usage of the optional Dock Cart
3-5-2-1
Attaching the Voluson i / Voluson e to the Dock Cart
1.) Hold the Voluson i / Voluson e slightly tilted towards you and place it on the top shelf next to
the arm rest.
2.) Now push the Voluson i / Voluson e away from the armrest until it locks in place.
3.) Push the probe connector into the probe socket of the Voluson i / Voluson e and lift the
locking lever upwards to lock the probe connector.The Probe Box enables the connection of up to
3 probes to the Voluson i / Voluson e.
4.) Connect the cables to the rear of the Voluson i / Voluson e. The cables to connect are listed
below. The order of connection is not important:
•
L-Shaped circular plug of the power cable
•
USB Cable - Connect the USB plug to the lower USB connector on the rear-panel of the
Voluson i / Voluson e. To connect USB Devices to the system, plug them into the Dock Cart’s
USB ports.
•
LAN cable - Connect the ethernet cable, routing from the LAN isolation box, into the network port
of the Voluson i / Voluson e.
5.) Place the DVD Drive on the Tray recorder (see
Figure 3-10 on page 3-14
) and connect it.
WARNING
WARNING
!! WARNING:
DO NOT CONNECT any devices to any AC power other than the on-board outlets from
the transformer, as this might increase leakage currents and sacrifice patient safety.
WARNING
WARNING
!! WARNING:
Before starting the Voluson i / Voluson e or Peripherals that are connected to the Dock
Cart, ALWAYS make sure that rating of the power supply fuses is compatible and the
voltage output selector is in the correct position.
The wrong fuses and position of the voltage output selector may cause major damage
on the Voluson i / Voluson e system and connected peripherals.
CAUTION
!! CAUTION:
Make sure that all peripherals are firmly secured in position using Velcro tapes or straps.
Place heavy and large devices on the lower storage shelf.
NOTICE
!! NOTICE:
A battery or an empty battery shell should always be attached to the Voluson i / Voluson e system
before attaching it to the Dock Cart.
Figure 3-28 Voluson i / Voluson e attached to the Dock Cart
probe connector
locking lever